![]() Inspiratory/Expiratory Ratio: 2:1 to 1:8 (increments of 0.5).Rate: 4 to 100 breaths per minute for Volume Control, Pressure Control and PCV-VG 2 to 60 breaths per minute for SIMV, PSVPro, SIMV PCV-VG 4 to 60 breaths per minute for CPAP+PSV (increments of 1 breath per minute).Pressure (P) Support Range: Off, 2 to 40 cmH2O (increments of 1 cmH2O).Pressure (P) Max Range: 12 to 100 cmH2O (increments of 1 cmH2O).Pressure (P) Inspired Range: 5 to 60 cmH2O (increments of 1 cmH2O) 5 to 1500 mL volume delivery.Minute Volume Range: Less than 0.1 to 99.9 L/min.Incremental Settings: 20 to 50 mL (increments of 1 mL) 50 to 100 mL (increments of 5 mL) 100 to 300 mL (increments of 10 mL) 300 to 1000 mL (increments of 25 mL) 1000 to 1500 mL (increments of 50 mL).Tidal Volume Range: 20 to 1500 mL (Volume Control, PCV-VG, SIMV and SIMV PCV-VG modes).Datex-Ohmeda device interface solutions port.Type: Optional galvanic fuel cell or paramagnetic with Respiratory Module option.Location: Inspiratory outlet and expiratory inlet.Volume Delivery: > 210 mL = better than 7% ≤ 210 mL = better than 15 mL 210 mL = better than 9% ≤ 210 mL = better than 18 mL Detection Through Machine Flow Transducers. ![]()